Frequently asked questions

Product use  |  Breathing circuits  |  Cleaning


Product use


Where is the F&P 850 System intended for use?

Critical care environments: intensive care, high dependency, emergency rooms, labor and delivery, neonatal intensive care units, special care baby units.


What is the humidity output?

The humidity performance in Invasive Mode is ≥ 33 mg/L. In Mask Mode, the humidity performance is ≥ 12 mg/L. In most cases, humidity delivery will be higher than these values, with the actual humidity delivery depending on environment conditions, respiratory gas properties and flow source settings.


What are the operating flow ranges?

At the recommended environmental conditions, the flow range for Invasive Mode is ≤ 60 L/min, and ≤ 120 L/min in Mask Mode. Refer to breathing circuit user instructions for the specific range.


How long does it take the humidifier to reach set dew point (with gas flow)?

The humidifier should take less than 30 minutes to reach the set dew point.




Breathing circuits


How are the F&P Evaqua 2 breathing circuits different from other RT-series circuits?

The inspiratory limb of F&P Evaqua™ 2 circuits has insulating MicroCell™ technology that reduces gas from cooling and forming condensate in the circuit. The expiratory limb has Evaqua 2 technology allowing water vapor to diffuse through the limb wall before it has an opportunity to condense into liquid water. These integrated technologies lead to less condensation compared with conventional heated humidifier circuits and reduce the need to break the circuit to empty condensate.


Can viruses and bacteria diffuse or permeate through the Evaqua 2 expiratory limb wall?

F&P Evaqua 2 incorporates technology that allows water vapor to diffuse through the material of the expiratory limb. Viruses and bacteria, including incredibly small virus particles, cannot permeate or diffuse through the material.


Do Fisher & Paykel Healthcare provide respiratory filters with F&P 850 breathing circuits?

The table below shows some of the F&P 850 circuits that include a respiratory filter.



  Circuit Kit

  Filter part number
  Filter position
  Filter description
  Duration of use
RT319RT016InspiratoryInspiratory/single limb use onlyUp to 24 hours

RT481

RT385

RT380

RT301

RT019Inspiratory & ExpiratoryInspiratory or end-expiratory use with an insulating filter housing to reduce condensationUp to 24 hours


Note: Inspiratory position represents the filter attached to the inspiratory port of the ventilator, and the end-expiratory position represents the filter attached to the expiratory port of the ventilator.

Not all filters and circuit kits are available in all countries. Consult your local Fisher & Paykel Healthcare representative for information on availability.




Cleaning


How should the humidifier be cleaned?

Using a damp cloth, clean the MR850 Humidifier with either isopropyl alcohol or normal dishwashing detergent. For accessory cleaning guidelines, refer to the User Instructions accompanying each accessory.

NOTE: Do not immerse the humidifier or accessory electrical connectors in any liquid.


How should the heater-wire adapter be cleaned?

Using a damp cloth, clean the heater-wire adapter with either isopropyl alcohol or normal dishwashing detergent.

NOTE: Do not immerse the heater-wire adapter in any liquid. This product is not designed for an autoclave sterilization process.


How should the temperature and flow probes be cleaned?

The F&P 850 temperature probes require reprocessing between patients. Refer to the temperature and flow probe user instructions for a list of cleaning options.